How to Apply the Decision Rule in a Calibration Results.

How to Apply the Decision Rule in a Calibration Results.

The following are sample data from the different areas of calibration.
The decision is based on the different rules to show a comparison.
1. Dimensional
Sample: Caliper

Existing Decision Rule (following APLAC TC 004)
Calibration Results
Specification/ Tolerance taken from JIS B 7507 : 2016
Instrument Error of External Measurement

Nominal

(mm)
Permissible Error

(mm)
Measured Value

(mm)
Error

(mm)
Decision
0
0.000.00Intolerance
20
20.02+0.02Intolerance
50± 0.0350.03+0.03Indeterminate
100
100.04+0.04Indeterminate
150
150.05**+0.05Out of tolerance

The expanded uncertainty is ± 0.01 mm, at approximately 95% confidence level with coverage factor, k = 2 


This item fulfils the requirements given in the above stated standard.

The requirements stated in this report are taken from the above stated standard.
Final Statement in reportMeasured value marked ** does not fulfil the requirements given in the above stated standard.

 

New: Simple Acceptance Rule (following ILAC G8)
Calibration Results
Specification/ Tolerance taken from JIS B 7507 : 2016.
Instrument Error of External Measurement

Nominal

(mm)
Permissible Error

(mm)
Measured Value

(mm)
Error

(mm)
Decision
0
0.000.00Intolerance
20
20.02+0.02Intolerance
50± 0.0350.03+0.03Intolerance
100
100.04**+0.04Out of tolerance
150
150.05**+0.05Out of tolerance

The expanded uncertainty is ± 0.01 mm, at approximately 95% confidence level with coverage factor, k = 2.


The measured values were observed in tolerance at the points calibrated.
Final Statement in reportMeasured value marked ** was observed out of tolerance at the points calibrated.



Samples for Simple Acceptance Decision Rule
The following are sample data from the different areas of calibration.
The decision is based on the different rules to show a comparison.
1. Dimensional
Sample: Micrometer
Existing Decision Rule (following APLAC TC 004)
Calibration Results
Specification/ Tolerance taken from BS 870 : 2008.

Feature InspectedAllowable Error
(inch)
Measured Value
(inch)
Expanded Uncertainty
(inch)
Decision

Flatness of Measuring Face

0.000050.00005± 0.00001Indeterminate
Parallelism of Measuring Face0.00010.00008± 0.00004Indeterminate
Range of Error of
Micrometer Screw
0.00010.00005± 0.00005Intolerance

The expanded uncertainty is at approximately 95% confidence level with coverage factor, k = 2.


This item fulfils the requirements given in the above stated standard.
Final Statement in

report
The requirements stated in this report are taken from the above stated standard.
 Measured value marked ** does not fulfil the requirements given in the above stated standard.


New: Simple Acceptance Rule (following ILAC G8)
Calibration Results
Specification/ Tolerance taken from BS 870 : 2008.

Feature InspectedAllowable Error(inch)Measured Value(inch)Expanded Uncertainty(inch)Decision
Flatness of Measuring Face0.000050.00005± 0.00001Intolerance
Parallelism of Measuring Face0.00010.00008± 0.00004Intolerance
Range of Error of
Micrometer Screw
0.00010.00005± 0.00005Intolerance

The expanded uncertainty is at approximately 95% confidence level with coverage factor, k = 2.

Final Statement in reportThe measured values were observed in tolerance at the points calibrated.

Measured value marked ** was observed out of tolerance at the points calibrated.

 

Samples for Simple Acceptance Decision Rule
The following are sample data from the different areas of calibration.
The decision is based on the different rules to show a comparison.
2. Mechanical
Sample: Pressure Gauge
Existing Decision Rule (following APLAC TC 004)
Calibration Results
Specification / Tolerance taken from Manufacturer's Specification : ± 0.1 % of Full Scale.

UUT Indicated ValueApplied PressureDecision
AscendingError 
00.000.00Intolerance
100100.02-0.02Intolerance 
300300.16-0.16Indeterminate
400400.4-0.41Indeterminate

All units are in psi.
The expanded uncertainty is ± 0.06 % of Full Scale, at approximately 95% confidence level with coverage factor, k = 2.


This item fulfils the requirements given in the above stated standard.
Final Statement in reportThe requirements stated in this report are taken from the above stated standard.

Measured value marked ** does not fulfil the requirements given in the above stated standard.

 

New: Simple Acceptance Rule (following ILAC G8)
Calibration Results
Specification / Tolerance taken from Manufacturer's Specification : ± 0.1 % of Full Scale.

UUT Indicated ValueApplied PressureDecision
AscendingError
00.000.00Intolerance
100100.02-0.02Intolerance
300300.16-0.16Intolerance
400400.41 **-0.41Out of tolerance

All units are in psi.
The expanded uncertainty is ± 0.06 % of Full Scale, at approximately 95% confidence level with coverage factor, k = 2.


The measured values were observed in tolerance at the points calibrated.
Final Statement in reportMeasured value marked ** was observed out of tolerance at the points calibrated.

 

Samples for Simple Acceptance Decision Rule
The following are sample data from the different areas of calibration.
The decision is based on the different rules to show a comparison.
2. Mechanical
Sample: Torque Wrench
Existing Decision Rule (following APLAC TC 004)
Calibration Results
Clockwise Torque

UUT Indicated Value(N.m)Max. Permissible
Relative Deviation(± %)
Mean Measured Value(N.m)Error expressed
as % of the Indicated Value
Expanded Uncertainty(%)Decision
5.0065.11-2.20±0.6Intolerance
15.00415.64-4.27±0.4Indeterminate
25.00425.98-3.92±0.4Indeterminate

Specification / Tolerance taken from ISO 6789-1 : 2017. Type I , Class B.
The expanded uncertainty is at approximately 95% confidence level with the reported coverage factor, k = 2.


This item fulfils the requirements given in the above stated standard.
Final Statement in reportThe requirements stated in this report are taken from the above stated standard.

Measured value marked ** does not fulfil the requirements given in the above stated standard.

 

New: Simple Acceptance Rule (following ILAC G8)
Calibration Results
Clockwise Torque

UUT Indicated Value(N.m)Max. Permissible
Relative Deviation(± %)
Mean Measured Value(N.m)Error expressed
as % of the Indicated Value
Expanded Uncertainty(%)Decision
5.0065.11-2.20±0.6Intolerance
15.00415.64**-4.27±0.4Out of tolerance
25.00425.98-3.92±0.4Intolerance

Specification / Tolerance taken from ISO 6789-1 : 2017. Type I , Class B.
The expanded uncertainty is at approximately 95% confidence level with the reported coverage factor, k = 2.


The measured values were observed in tolerance at the points calibrated.
Final Statement in reportMeasured value marked ** was observed out of tolerance at the points calibrated.

 

Samples for Simple Acceptance Decision Rule
The following are sample data from the different areas of calibration.
The decision is based on the different rules to show a comparison.
3. Electrical
Sample: Multimeter
Existing Decision Rule (following APLAC TC 004)
Calibration Results

Performance TestInstrument ReadingManufacturer's SpecExpanded Uncertainty(±)Decision
Input Freq(Hz)
Lower LimitUpper Limint
 
Ac Voltage Test  



 
6.0000.0456.5 mV5.6 mV6.40.1Indeterminate
600.000.01K599.9 mV587.7 mV612.30.1Intolerance
5.0000.0455.052 V4.947 V5.0530.002Indeterminate



This item fulfils the requirements given in the above stated standard.
Final Statement in reportThe requirements stated in this report are taken from the above stated standard.

Measured value marked ** does not fulfil the requirements given in the above stated standard.

 

New: Simple Acceptance Rule (following ILAC G8)
Calibration Results

Performance TestInstrument ReadingManufacturer's SpecExpanded Uncertainty(±)Decision
Input
Freq(Hz)
Lower LimitUpper Limint









6.0000.0456.5** mV5.6 mV6.40.1Out of tolerance
600.000.01K599.9 mV587.7 mV612.30.1Intolerance 
5.0000.0455.052 V4.947 V5.0530.002Intolerance 



The measured values were observed in tolerance at the points calibrated.
Final Statement in reportMeasured value marked ** was observed out of tolerance at the points calibrated.

 

Samples for Simple Acceptance Decision Rule
The following are sample data from the different areas of calibration.
The decision is based on the different rules to show a comparison.
4. Temperature
Sample: Thermohygrograph
Existing Decision Rule (following APLAC TC 004)
Calibration Results
Manufacturer's Specification : ± 1°C and ± 5%rh

Reference ValuesUUT ValuesDeviationDecision
°C%rh°C%rh°C%rh
23.255.32356-0.20.7Indeterminate
27.365.428.4651.1-0.4Indeterminate
  1. The calibration results reported indicate all the test results obtained at the time of testing.
  2. The expanded uncertainty is ± 1.0°C and ± 5%rh, at approximately 95% confidence level with coverage factor, k = 2.

The measured values were observed in tolerance at the points calibrated.
Final Statement in reportMeasured value marked ** was observed out of tolerance at the points calibrated.