Using decision rules to analyze calibration results

Using decision rules to analyze calibration results

Calibration is an important process to ensure the accuracy and reliability of various measuring devices. Calibration allows the measurement devices to be checked and improved so that they can perform as intended. In analyzing calibration results, decision rules are an important tool to help decide whether the calibration results pass the acceptance criteria. This article will explain how to use decision rules to obtain accurate results.


Decision rules are the rules that determine how to interpret the results of calibration and decide whether the calibrated device is acceptable or not. Decision rules can be divided into several types, depending on the nature of the measurement and the requirements of the device. For example,

1. Simple Decision Rule: Used when there is a single acceptance criteria, such as if the measured value is within a specified range, the device is considered calibrated.
2. Conditional Decision Rule: Used when there are multiple acceptance criteria or additional conditions, such as if the measured value exceeds the specified criteria, the device may need to be recalibrated or improved.

Steps to Use Decision Rules in Analyzing Calibration Results
1. Define Acceptance Criteria
Defining acceptance criteria is an important first step in using decision rules. The specified value is the value that the measuring device must be within the specified range in order to be considered a calibration result that passes the criteria.
  • Example: For the calibration of a thermometer, the acceptance criteria may be set at ±1°C from the standard value. If the measured temperature value falls within this range, the calibration is considered to have passed.
2. Uncertainty Evaluation
The assessment of uncertainty is an important step to know the uncertainty that occurs in the calibration process. The calculated uncertainty will help to better assess the accuracy of the calibration result.
  • Example: For the calibration of a thermometer, it may be necessary to calculate the uncertainty from repeated measurements and review the standard value, such as an uncertainty of ±0.5°C.
3. Comparison with the acceptance criteria
This step is to compare the calibration result with the specified acceptance criteria.
  • Example: If the temperature value measured by the measuring device is within the range of 99°C to 101°C (100°C ± 1°C), the calibration is considered to have passed.
4. Decision based on the results
After comparing the calibration results with the acceptance criteria, a decision must be made whether the calibration result passes or not.
  • Example: If the calibration result passes the acceptance criteria, the device can be used normally. If the calibration result does not pass the acceptance criteria, It is necessary to adjust or re-calibrate until the results pass the criteria.


Example of using the decision rule
Suppose we have a blood pressure monitor calibration, which has an acceptance criteria of ±5 mmHg from the standard value, with the following steps:

1. Measuring the standard value
  • Measure the blood pressure value from a standard source, such as a known pressure (120/80 mmHg)
2. Measuring the value from the measuring device
  • Use the blood pressure monitor to be calibrated to measure the same pressure value as the standard value
3. Uncertainty assessment
  • Calculate the uncertainty of the measurement, such as an uncertainty of ±2 mmHg
4. Comparison with the acceptance criteria
  • If the blood pressure value measured by the device is in the range of 115/75 mmHg to 125/85 mmHg (120/80 mmHg ± 5 mmHg), it is considered calibrated.
5. Decision making
  • If the measured blood pressure value does not exceed the criteria of ±5 mmHg, the device has passed the calibration and can be used. If the measured blood pressure value exceeds the ±5 mmHg threshold, adjustment and recalibration must be performed.


The importance of using decision rules
Using decision rules to analyze calibration results is very important because:
  • Increases accuracy and reliability: Using decision rules helps make calibration more accurate and reliable, allowing for accurate and efficient decisions.
  • Reduces risk: Using decision rules reduces the risk that may arise from using uncalibrated instruments, which may cause measurement errors.
  • Improves quality: Effective calibration allows for continuous improvement of the quality of measurement devices and processes.
Conclusion
Using decision rules to analyze calibration results is a method that helps make calibration more accurate and reliable. By setting acceptance criteria, evaluating uncertainty, and making decisions based on the results, it is possible to ensure that calibrated devices can be used as intended.